Baumer IVO GmbH & Co. KG 02.10 • 171.02.281/8 Dauchinger Strasse 58-62 • DE-78056 Villingen-Schwenningen Subject to modification Phone +49 (0)77
N 150 www.baumer.com 10 4.2.8. Indicate optional column of figures in bottom line „u“ (75h) The command is utilized to indicate a 5-digit column o
N 150 www.baumer.com 11 4.3. Parameter commands Command code Data amountin bytes read write Broad- cast Save in EEPROM Significance a (61h) 5
N 150 www.baumer.com 12 Significance of parameter „a“ Positioning direction Positioning direction This parameter defines the direction the target
N 150 www.baumer.com 13 Resolution This parameter defines the resolution of the actual value. Upon altering the resolution the decimal point is sh
N 150 www.baumer.com 14 4.3.4. Read / write measuring unit „i“ (69h) The command is utilized to read or write the measuring unit in mm or inches.
N 150 www.baumer.com 15 4.4. Identifier commands Command code Data amountin bytes read write Broad- cast Save in EEPROM Significance A (41h) 2
N 150 www.baumer.com 16 Extended identifier command „AX“ (41h,58h) This extended command is utilized to address individual SPAs in the bus network
N 150 www.baumer.com 17 4.5. Specific commands Command code Data amountin bytes read write Broad- cast Save in EEPROM Significance K (4Bh) -
N 150 www.baumer.com 18 4.5.3. Read version, type or serial number „X“ (58h) This command is utilized to read version number, device type or seria
N 150 www.baumer.com 19 5. Error warnings 5.1. CRC error Upon recognizing a CRC error in a transmitted command the SPA will respond as follows:
N 150 www.baumer.com 2 1. General information 1.1. Safety precautions General remarks The equipment is designed and assembled according to the
N 150 www.baumer.com 20 7. Technical data Technical data - electrical ratings Supply voltage 24 VDC ±10 % Current consumption ≤30 mA Display
N 150 www.baumer.com 21 7.1. Dimensions 601352Aø12ø12A30015010M8femaleconnectorM8connectorCable outputConnect or output7520.5221613376.5451430.153
N 150 www.baumer.com 3 1.2. SPA description and explanation Mounting the spindle position display (SPA) is by docking hollow shaft onto spindle s
N 150 www.baumer.com 4 2. Terminal assignment Pin Assignment Core colour DIN47100 IEC757 Pin 1 Tx/Rx-, RS485 white orange Pin 2 Tx/Rx+, RS48
N 150 www.baumer.com 5 3. Interface 3.1. Interface data Type: RS485 Baudrate: 19200 Parity: No Data bits: 8 Stop bits: 1 Handshake: No Checksum:
N 150 www.baumer.com 6 3.4. Data transmission properties For reasons of compatibility, within the complete multicon system the data amount is alw
N 150 www.baumer.com 7 4.2.1. Check Position „C“ (43h) This command provides the alignment status of target versus actual value. If the actual val
N 150 www.baumer.com 8 4.2.4. Read /write offset „U“ (55h) This command is utilized to read or write the offset that is added to the true actual
N 150 www.baumer.com 9 4.2.6. Set actual value as preset „Z“ (5Ah) Command “Z” is utilized to set the actual value at any optional value. The req